A systematic quality assurance study in bone densitometry devices


Tuncman D., Kovan H., Kovan B., Demir B., Turkmen C.

International Workshop on Theoretical and Experimental Studies in Nuclear Applications and Technology (TESNAT), Osmaniye, Türkiye, 23 - 26 Nisan 2015, cilt.100 identifier identifier

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Cilt numarası: 100
  • Doi Numarası: 10.1051/epjconf/201510003004
  • Basıldığı Şehir: Osmaniye
  • Basıldığı Ülke: Türkiye
  • İstanbul Üniversitesi Adresli: Evet

Özet

Osteoporosis is the most common metabolic bone disease and can result in devastating physical, psychosocial, and economic consequences. It occurs in women after menopause and affects most elderly. Dual-energy x-ray absorptiometry (DXA) is currently the most widely used method for the measurement of areal Bone Mineral Density (BMD) (g/cm(2)).DXA is based on the variable absorption of X-ray by the different body components and uses high and low energy X-ray photons. There are two important values in the assessment of the DXA. These values are T-score and Z-score. The T-score is calculated by taking the difference between a patient's measured BMD with the mean BMD of the young normal population, matched for gender and ethnicity, and then by dividing the difference with the standard deviation (SD) of the BMD of the young normal population. T-score and also Z-score are directly depends on the Bone Mineral Density (BMD). BMD measurements should be made periodically in a patient life. But mostly, it is not possible with the same device. Therefore, in this study, for the quality assurance of bone densitometry devices, we evaluated the BMD results measured in the different Bone Densitometry (DXA) devices using a spine phantom.