A hybrid method to determine efficiency curve of HPGe detectors


Özben C. Ş., Emirhan E. M.

APPLIED RADIATION AND ISOTOPES, cilt.67, sa.6, ss.1110-1113, 2009 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 67 Sayı: 6
  • Basım Tarihi: 2009
  • Doi Numarası: 10.1016/j.apradiso.2009.01.085
  • Dergi Adı: APPLIED RADIATION AND ISOTOPES
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.1110-1113
  • İstanbul Üniversitesi Adresli: Hayır

Özet

A method was developed for determining the efficiency calibration curve of HPGe detectors between 26 and 1770 keV. A simple simulation based on the attenuation of gamma-rays in pure germanium was used together with efficiency calibrations performed for three detector-source distances: 5, 12 and 30 cm. The simple simulation model required an energy dependent term with three parameters and these parameters were determined from the fit of the simulated data to the measured efficiency of the detector. The efficiency response of the detector determined this way was then tested with other certified sources to verify the validity of the method. (C) 2009 Elsevier Ltd. All rights reserved.