The effects of kilovoltage (kV) and milliampere seconds (mAs) values for SPECT attenuation correction: An anthropomorphic phantom study


Kovan B.

Applied Radiation and Isotopes, cilt.204, 2024 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 204
  • Basım Tarihi: 2024
  • Doi Numarası: 10.1016/j.apradiso.2023.111118
  • Dergi Adı: Applied Radiation and Isotopes
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Academic Search Premier, Aquatic Science & Fisheries Abstracts (ASFA), Chemical Abstracts Core, Chimica, Compendex, EMBASE, Food Science & Technology Abstracts, INSPEC, MEDLINE, Pollution Abstracts
  • İstanbul Üniversitesi Adresli: Evet

Özet

The purpose of SPECT/CT imaging to be used in dosimetric calculations is to perform accurate attenuation correction by giving the patient the minimum radiation dose. in accordance with ALARA principles (as low as reasonably achievable), it is critical to limit the amount of radiation given to the patient as far as possible in medical fields where radiation therapy is applied. The goal of this study is to determine the optimum kV and mAs values needed to keep the radiation dose that can be given to the patient in CT scans at a minimum level for attenuation correction, as well as to determine the CT-induced radiation doses on patients with different kV and mAs values using an anthropomorphic phantom. Radionuclides 177Lu and 131I were used in SPECT imaging. For CT-based attenuation correction, images were taken with various values between 80 kV × 30 mAs and 140 kV × 300 mAs using the anthropomorphic phantom. After that attenuation correction was performed with the obtained CT images for SPECT images. Then, CT-derived radiation doses were calculated. Our results showed that 80 kV × 30 mAs values can be used in scans after radionuclide treatment for CT attenuation correction. It was determined that the patient was exposed to a radiation dose of 0.432 mSv as a result of the CT scanning with the full detector width area of gamma camera detector. Thus, scanning with these values (80 kV × 30 mAs) resulted in approximately 95% less radiation exposure (0,432 mSv) than the value (9,38 mSv) with scanning of the manufacturer's recommended value (120 kV × 200 mAs).