Rupture of the NovaFlex balloon during transcatheter aortic valve implantation and subsequent dissection of the right iliac arteries


Gul M., Aksu H. U., Oz K., BAKIR İ.

EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, cilt.43, sa.2, ss.437-438, 2013 (SCI-Expanded) identifier

  • Yayın Türü: Makale / Editöre Mektup
  • Cilt numarası: 43 Sayı: 2
  • Basım Tarihi: 2013
  • Doi Numarası: 10.1093/ejcts/ezs508
  • Dergi Adı: EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.437-438
  • Anahtar Kelimeler: TAVI, Complication, Balloon rupture, FOLLOW-UP
  • İstanbul Üniversitesi Adresli: Evet

Özet

We report a case with a very rare complication of transcatheter aortic valve implantation. Rupture of the NovaFlex balloon (Edwards transfemoral balloon catheter) occurred during the inflation of the Edwards SAPIEN valve, resulting in dissection of the right common and external iliac arteries during withdrawal of the balloon catheter. The NovaFlex balloon is a part of the Edwards NovaFlex XT transfemoral delivery system.